PERBEDAAN NILAI IMAGE NOISE DAN DOSIS RADIASI DENGAN MENGGUNAKAN AUTOMATIC EXPOSURE CONTROL (AEC) PADA PEMERIKSAAN CT SCAN

Authors

  • Lutfatul Fitriana Universitas Muhammadiyah Purwokerto
  • Hernastiti Sedya Utami Universitas Muhammadiyah Purwokerto

DOI:

https://doi.org/10.38165/jk.v12i2.259

Abstract

Dosis radiasi merupakan hal yang perlu diperhatikan dalam pemeriksaan CT scan. Pada pesawat CT scan terdapat software automatic exposure control yang berfungsi menentukan nilai mAs secara otomatis yang disesuaikan dengan ketebalan objek, sehingga diharapkan dapat mengurangi dosis radiasi yang diterima oleh objek. Penelitian ekperimen ini menggunakan 3 variasi kV yaitu 80 kV, 110 kV dan 130 kV, masing-masing nilai kV diberi perlakuan menggunakan automatic exposure control dan tanpa menggunakan automatic exposure control. Hasil penelitian yang didapat yaitu ada perbedaan nilai dosis radiasi dan image noise anatar penggunaan automatic exposure control dan tanpa penggunaan automatic exposure control, dimana penggunaan automatic exposure control dosis radiasi yang dihasilkan jauh lebih rendah dibandingkan tanpa penggunaan automatic exposure control. Sedangkan untuk nilai image noisenya, penggunaan automatic exposure control menghasilkan nilai image noise lebih tinggi dibandingkan tanpa penggunaan automatic exposure control. Sehingga penggunaan automatic exposure control ini dapat menurunkan dosis radiasi yang diterima objek tetapi juga meningkatkan image noise

Kata kunci: Image Noise, Dosis Radiasi, AEC

 

 

Abstract

Radiation dose is something that needs to be considered in a CT scan. On the CT Scan aircraft there is automatic exposure control software that functions to determine the mAs value automatically which is adjusted to the thickness of the object, so that it is expected to reduce the radiation dose received by the object. This experimental study uses 3 variations of kV, namely 80 kV, 110 kV and 130 kV , each kV value was treated using automatic exposure control and without using automatic exposure control. The results obtained are that there are differences in the value of radiation dose and image noise between the use of automatic exposure control and without the use of automatic exposure control, where the use of automatic exposure control produces a much lower radiation dose than without the use of automatic exposure control. As for the image noise value, the use of automatic exposure control results in a higher image noise value than without the use of automatic exposure control. So that the use of automatic exposure control can reduce the radiation dose received by the object but also increase image noise

Keywords: Image Noise, Radiation Dose, AEC

Author Biography

Hernastiti Sedya Utami, Universitas Muhammadiyah Purwokerto

Program Studi D4 Teknologi Radiologi Pencitraan

Downloads

Published

2021-12-29

How to Cite

Fitriana, L., & Utami, H. S. (2021). PERBEDAAN NILAI IMAGE NOISE DAN DOSIS RADIASI DENGAN MENGGUNAKAN AUTOMATIC EXPOSURE CONTROL (AEC) PADA PEMERIKSAAN CT SCAN. Jurnal Kesehatan, 12(2), 131–136. https://doi.org/10.38165/jk.v12i2.259

Issue

Section

Articles

Most read articles by the same author(s)